Mar 25, 2019Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior acceCVE-2015-3956 · Other software · Hospira · reported by ics-cert@hq.dhs.govCritical 9.8
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
- Product
- Plum A+ Infusion System
- What to do
- Apply the vendor’s update. The NVD record links to the advisory.
Full record and what to do →NVD recordics-cert.us-cert.govics-cert.us-cert.gov
Mar 25, 2019Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior giveCVE-2015-3954 · Other software · Hospira · reported by ics-cert@hq.dhs.govCritical 9.8
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
- Product
- Plum A+ Infusion System
- What to do
- Apply the vendor’s update. The NVD record links to the advisory.
Full record and what to do →NVD recordics-cert.us-cert.govics-cert.us-cert.gov
Mar 25, 2019Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq InCVE-2015-3953 · Other software · Hospira · reported by ics-cert@hq.dhs.govCritical 9.8
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
- Product
- Plum A+ Infusion System
- What to do
- Apply the vendor’s update. The NVD record links to the advisory.
Full record and what to do →NVD recordics-cert.us-cert.govics-cert.us-cert.gov
Mar 25, 2019Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq InCVE-2015-3952 · Other software · Hospira · reported by ics-cert@hq.dhs.govHigh 7.5
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
- Product
- Plum A+ Infusion System
- What to do
- Apply the vendor’s update. The NVD record links to the advisory.
Full record and what to do →NVD recordics-cert.us-cert.govics-cert.us-cert.gov
Jan 22, 2016Stack-based buffer overflow in Hospira Communication Engine (CE) before 1.2 in LifeCare PCA Infusion System 5.07, Plum A+ Infusion System 13.40, and Plum A+3 InCVE-2015-7909 · Other software · Hospira · reported by ics-cert@hq.dhs.govHigh 7.3
Stack-based buffer overflow in Hospira Communication Engine (CE) before 1.2 in LifeCare PCA Infusion System 5.07, Plum A+ Infusion System 13.40, and Plum A+3 Infusion System 13.40 allows remote attackers to cause a denial of service or possibly have unspecified other impact via traffic on TCP port 5000.
- Product
- Communication Engine
- What to do
- Apply the vendor’s update. The NVD record links to the advisory.
Full record and what to do →NVD recordics-cert.us-cert.govics-cert.us-cert.gov
Where this comes from: the exploited list is CISA’s Known Exploited Vulnerabilities catalog. Everything else is the whole National Vulnerability Database, where each record is published by the team that found the bug (Wordfence, Patchstack, WPScan, the vendors themselves and others); we show who. Exploit odds are FIRST’s EPSS, updated daily. Titles and categories are ours, so check the linked record before acting on a detail. For a write-up of the big ones, see our advisories. Every vendor, and every WordPress plugin and theme with a recorded flaw, also has its own page. All of it is free to reuse as JSON, CSV or RSS.
Weekly security brief
The flaws that hit local websites, every Friday.
One short email: what got exploited this week, which WordPress plugins broke, and what to check. Unsubscribe in one click.
Tell us what you're building.
A website, your marketing, a security check, an app, or the thing you've been running out of a spreadsheet for three years. A few lines is plenty.
Not ready to talk yet? Run the free website check and see what Google sees.